Material: | Stainless Steel |
---|---|
Type: | Biosafety Cabinet |
Usage: | Hospital, Lab |
Still deciding? Get samples of US$ 2700/unit
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Advantage:
1. Time reserve function.
2. Motorized front window.
3. ULPA filter life and UV life indicator.
4. Automatic air speed adjustable with filter block.
5. With memory function in case of power-failure.
6. Work area surrounded by negative pressure, it can ensure maximum safety in work area.
7. Most accessories are standard, which saves your money. There is no need to pay more.
8. Audio and visual alarm (filter replacement, window over height, abnormal air flow velocity, etc).
9. Remote control. All functions can be realized with it, making the operation much easier and more convenient.
10. Interlock function: UV lamp and front window; UV lamp and blower, fluorescent lamp; blower and front window.
11. Foot switch. Adjust front window height by foot during experiment, to avoid airflow turbulence caused by arm movement.
Model |
BSC-1100IIB2-X |
BSC-1300IIB2-X |
BSC-1500IIB2-X |
BSC-1800IIB2-X |
|
External Size (W*D*H) |
1100*750*2250mm |
1300*750*2250mm |
1500*750*2250mm |
1800*775*2250mm |
|
940*600*660 mm |
1150*600*660 mm |
1350*600*660mm |
1625*600*660 mm |
||
Tested Opening |
Internal Size (W*D*H) Safety Height 200 mm(8'') |
||||
Max Opening |
420mm(17'') |
420mm(17'') |
500mm(20'') |
480mm(20'') |
|
Inflow Velocity |
0.53±0.025 m/s |
||||
Down flow Velocity |
0.33±0.025 m/s |
||||
ULPA Filter |
Two, 99.9995% efficiency at 0.12 μm,filter life indicator. |
||||
Front Window |
Motorized,two-layer laminated toughened glass ≥ 6mm,anti UV. |
||||
Noise |
≤ 67 dB |
||||
UV Lamp |
30W*1 |
30W* 1 |
40W* 1 |
40W*1 |
|
UV timer, UV life indicator, emission of 253.7 nanometers for most efficient decontamination. |
|||||
Illuminating Lamp |
LED Lamp |
||||
12W*2 |
12W*2 |
16W*2 |
16W*2 |
||
Illumination |
≥1000Lux |
||||
Consumption |
800W |
850W |
1300W |
1300W |
|
Waterproof Sockets |
Two, total load of two sockets: 500W |
||||
Display |
LCD display: exhaust filter and downflow filter pressure, filter and UV lamp working time, inflow and downflow velocity, filter life, humidity and temperature, system working time etc. |
||||
Control System |
Microprocessor |
||||
Airflow System |
0% air recirculation, 100% air exhaust |
||||
Alarm |
Abnormal airflow velocity; Filter replacement; Front window at unsafe height. |
||||
Exhaust Duct |
4 meters PVC duct, Diameter:300mm |
||||
Material |
Work Zone: 304 stainless steel |
||||
Main Body: Cold-rolled steel with anti-bacteria powder coating. |
|||||
Work Surface Height |
750mm |
||||
Caster |
Foot master caster |
||||
Power Supply |
AC 220V±10%, 50/60Hz; 110V±10%, 60Hz (110V/60Hz is not applicable to BSC-1800IIB2-X) |
||||
Standard Accessory |
Illuminating lamp, UV lamp*1,base stand, remote control, foot switch, exhaust blower, exhaust duct, drain valve, waterproof sockets*2, pipe strap*2 |
||||
Optional Accessory |
Water and gas tap, Electric height adjustable base stand |
||||
Gross Weight |
240kg |
280kg |
300kg |
408kg |
|
Package |
Main Body |
1220* 1000* 1820 mm |
1460*1000*1820 mm |
1630*1000*1820 mm |
1930*1000*1820 mm |
Exhaust Blower (W*D*H) |
970* 810* 630 mm |
970* 810* 630 mm |
970* 810* 630 mm |
970* 810* 680 mm |