After-sales Service: | Online After Sales Service |
---|---|
Warranty: | 1 Year |
Appliance: | Food Analysis |
Still deciding? Get samples of US$ 5600/unit
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Features:
*Affordable
*VFD display screen
*Control mode: external/internal
Model
|
Agress1100(Type I)
|
Agress1100(Type II)
|
BK-LCG1100 |
|
High Pressure Constant Flow Pump
|
Double plunger reciprocating pump
|
1 set
|
2 sets
|
2 sets |
Flow Range
|
0.001mL/min-10.000mL/min
|
|||
Setting Step Size
|
0.001mL/min
|
|||
Flow Accuracy
|
≤±0.3% (1.00ml/min, H2O,R.T.,8.5MPa)
|
|||
Flow stability
|
RSD≤0.1% (1.00ml/min, H2O,R.T.,8.5MPa)
|
|||
Maximum Pressure
|
42MPa
|
|||
Display Pressure Error
|
≤±3.0%
|
|||
Pressure Fluctuation
|
≤1.0%
|
|||
Power Dissipation
|
80W
|
|||
UV/VIS Detector |
Wavelength Range
|
190-700nm
|
||
Wavelength Repeatability
|
±0.1nm
|
|||
Wavelength Accuracy
|
≤±1.0nm
|
|||
Spectrum Band Width
|
8nm
|
|||
Response Time
|
0.0s-4.9s
|
|||
Linearity Range
|
1.8AU (5%)
|
|||
Baseline Noise
|
≤±1×10-5AU
|
|||
Baseline Drife
|
≤2.0×10-4AU/h
|
|||
minimum Detectable Concentration
|
4.0×10-9g/mL(Naphthalene in Methanol Solution)
|
|||
Detector Cell
|
Optical Path--10mm, Volume--10μL
|
|||
Power Dissipation
|
100W
|
|||
Gradient Mixer
|
/
|
GM1100
|
GM1100 |
|
Sampler Injector
|
1
|
|||
Supersil ODS2 5μm ID4.6×200mm
|
1
|
|||
Workstation (Without Computer & Printer)
|
1
|
|||
Valve Mounting Frame
|
1
|
|||
Kits
|
Isocratic System Kits
|
Gradient System Kits
|
Gradient System Kits |
|
Power Supply
|
AC110/220V±10%, 50/60Hz
|
|||
Optional Accessory
|
Column Oven
|