BK-1200 Auto Chemistry Analyzer
Features:
1. 150 Sample positions
2. 160 Reaction cuvettes
3. 180 Refrigerated reagent positions
4. Optical filter/fiber bragg grating selectable
5. 800 Tests per hour 1200 Tests with ISE module
Overall
Performance |
Model
|
BK-800(EMERALD)
|
Type
|
Random, Discrete, Fully automatic
|
|
Test Speed
|
800T/H, 1200T/H with ISE
|
|
Analysis Method
|
End-point, Fixed-time, Rate, Turbidimetry
|
|
Programming Items
|
Programmable for any test items
|
|
Sample unit
|
Sample Position
|
150 sample positions, including 60 refrigerated positions for calibrator&QC, dual tray
|
Sample Tubes
|
Original blood collection tube, micro tube, plastic tube, etc
|
|
Sample Types
|
Serum, plasma, urine, CFS, etc
|
|
Sample Volume
|
2μl~70μl (0.1μl stepping)
|
|
Sample Probe
|
Liquid level sensor function, anti-collision function
|
|
Probe Cleaning
|
Automatically wash interior and exterior, cross contamination rate <0.1%
|
|
Analysis Items
|
Analyzing maximally 100 items at the same time
|
|
Reagent unit
|
Reagent Position
|
180 reagent positions, dual tray
|
Wash Solution
|
Washing solution can be installed in any position, including needle, stir bar
washing solution and water bath disinfectant |
|
Reagent Volume
|
20~350μl (0.5μl stepping)
|
|
Temperature Control
|
2~8ºCcooling for 24 hours
|
|
Reagent Probe
|
Liquid level sensor function, anti-collision function
|
|
Probe Washing
|
Automatically wash interior and exterior
|
|
Analysis Items
|
Analyzing maximally 100 items at the same time
|
|
Reaction unit
|
|
|
|
Temperature System
|
|
Cuvettes
|
160 positions, PMMA cuvette (optical path length 5mm)
|
|
Mixer Probe
|
Independent stir bar, nanotechnology, highly polished, automatic frequency conversion
|
|
Washing
|
Automatic water cleaning and drying system
|
|
System Adjustment
|
System can record the blank value and get more accurate
result by deducting blank base value of each cuvette |
|
Optical System
|
Light Source
|
Halogen lamp (Grating optional)
|
Receiver
|
Holographic flat field concave grating diode array detection
|
|
Wavelength Range
|
12 wavelengths from 340 to 800nm
|
|
Linear Range
|
-0.3~3.3 Abs
|
|
Stray Light
|
<0.1% or Abs >3.0
|
|
Calibration
and QC |
Calibration Method
|
Linear calibration: K factor, two-and multiple linear
Nonlinear calibration: spline, polynomial, logarithm, etc |
Quality Control
|
Auto QC, westgard, cumulative sumcheck, twinplot
|
|
Software
|
STAT Function
|
Emergency samples can be placed in the sample tray at any time
|
Computer System
|
Windows 7/8/10, LIS system available
|
|
Online
|
Full bi-directional communication possible
|
|
Data Processing
|
Reaction curve and data display function
|
|
Others
|
Water consumption
|
Deionized water<25L/H
|
Power Supply
|
AC110/220V±10%, 60/50Hz
|
|
Package Size
|
1718*929*1434mm
|
|
Gross Weight
|
340kg
|
Packaging
-Cartons packaging
-Customized packaging can be available.
Shipping
- International Express such as DHL, TNT, EMS, FedEX and so on .
- Sea Transportation.
-Air Transportation.